2.
6 / 108
1.

General Information

1.1. Weather and environmental conditions

The SWG100BIOGAS analyser is designed for operating ambient temperatures of +5°C to
+45°C (without cabinet heater) resp. -20°C to +45°C (with cabinet heater).
The SWG100BIO-Ex. is for designed for use in hazardous Zone 2, see the conformity marking
in the corresponding section.
In case of outdoor mounting it is important that the analyser is protected against rain and sun.
For extreme environmental conditions like high humidity, salty sea air, etc. further protective
measures are necessary. These should be clarified with the manufacturer (MRU).
Any additional protective measures for outdoor mounting have to be provided by the plant
operator. The manufacturer (MRU) is supporting the plant operator in choosing appropriate
protective measures.

1.2. Installation instructions

Installation instructions, which are described in chapter 17 of the operation manual, have to be
strictly carried out.
Please check correct installation before powering up the analyzer!
Please operate the analyzer only in upright position!

1.3. General important instructions for the plant operator

To guarantee continuous operation of the SWG100 BIOGAS analyser, the functions, processes
and operation of the analyser have to be monitored regularly by the plant operator – especially
in case of any initial installation. Thus, it will be possible to take suitable measures to improve
the availability and life time of the analyser.
As the plant operator gains more experience concerning the maintenance requirements of the
analyser, the monitoring frequency may be reduced to more extended periods of time.
It is important that the filter-units, listed in chapter 10 of the operation manual, are checked
regularly and, if necessary, service parts are changed.
In case of not intended use the guarantee will void. Regular controls,
inspections and the exchange from polluted and exhausted filters by
the operator are also an important part of the determination "not
determined use". See chapter 9 regular maintenance work.
SWG100 BIOGAS User Manual
WARNING
ATTENTION
MRU GmbH, D-74172 Neckarsulm