Ensure Safety; Important General Information En50379 And Vdi4206 - mru OPTIMA7 biogas User Manual

Hide thumbs Also See for OPTIMA7 biogas:
Table of Contents

Advertisement

1.3. Ensure safety

► Please read the user manual completely before the first use.
► Only use the analyser for the intended use and within the parameters
specified in the technical data.
► Do not use any violence.
► Avoid falls.
► Do not put the analyser into use, if the housing, power supply unit or
supply leads are damaged.
► Do not store the analyser together with solvents. Do not use
desiccants.
► Only carry out maintenance and service work on this analyser as de-
scribed in the User Manual. Observe the prescribed action steps.
► Operate the analyser only in closed, dry rooms and protect it from
rain and moisture.
► When operating the analyser on mains power, operate it only with
the mains adapter supplied.
► Do not use the metal tube of the gas sampling probe or other metal-
lic parts / accessories as electrical conductors.
► The analyser must not be located in the immediate vicinity of open
fire or great heat.
► The specified temperature range of the gas sampling probe must not
be exceeded, otherwise the probe tube and temperature sensor will
be destroyed.

1.4. Important general information EN50379 and VDI4206

This analyser is not designed to be used for continuous measurements.
Before using the analyser verify the condition of the various parts of the
analyser, such as the probe, the ambient air conditions, the condensate
separator, star filter and the connectors for damage and/or blockages.
When starting up the analyser it will take between 1 – 3 minutes to set to
zero depending on the condition of the sensors and of ambient.
The minimum zeroing time of the analyser to achieve correct measure-
ment values can be expected by 1.5 minutes.
Exposure to acids; aggressive gases such as sulphur; vapours such as
thinners, gasoline, alcohol and paint, etc. can damage, reduce the life of,
or destroy the sensors.
The life of the sensors depends on how they are used, maintained and
treated. Typical average life expectations are: O
H
S 2 years and NO - 3 years.
2
The use of the analyser for regulatory purposes is subject to special reg-
ulations (for example a periodical examination of the analyser). Please
obtain the appropriate regulations from your local responsible authority.
MRU GmbH, D-74172 Neckarsulm
USER MANUAL OPTIMA 7 Biogas
- 3 years; CO - 2 - 3 years,
2
7 / 78