Leave the menu and safe the adjustments. The individual configuration is done.
16.2.
The instrument may be optionally equipped with a H
limited life time when experiencing high H2S concentrations, some precautions are taken to
protect the sensor:
32 / 108
5. Leave the cycle phase detail and select the next phase. Select with the
arrow keys
the phase "zeroing" and go in the cycle phase detail. Adjust the phase-
duration at 15 min. and leave the cycle phase detail.
6. Switch to the next point and select with the
SP1. Go in the cycle phase detail of the measurement SP1. Here adjust the duration-
phase 10 min. and the pure measurement time: 5 min.
7. At last push
F3
key for insert a new phase and select with the
"Stand-by" phase. Go into the cycle phase detail and adjust the Purging time at 3 min.
and the Quiet-time at 10 min.
H
S
sensor protection (optional)
2
low
The instrument is equipped with a cut-off solenoid valve and air purge pump to protect
•
the H
S
sensor, without impacting the results of all other sensors.
2
low
The purge unit is always activated after 3-10 minutes of active sensor measuring time.
•
left/right arrow key
Or
F1/F3
Or
F1/F3
S
sensor. As this sensor typically shows
2
low
SWG100 BIOGAS User Manual
left/right
the measurement
left/right arrow
key the
Or
F1/F3
MRU GmbH, D-74172 Neckarsulm