Technical data

WATER SUPPLY SYSTEM (TO BE COMPLIANT WITH IEC61770)
Fitting
Backflow preventing device (IEC61770).
Temperature
max. 35 °C (95 °F)
Pressure
min. 2 bar – max. 8.6 bar (min 29 psi - max. 124.7 psi)
Flow
min. 0.25 – max. 0.5 l/min (min. 0.066 - max. 0.132 gal/min)
POWER SUPPLY SYSTEM
Nominal voltage and Max.
200–240 V ac, 50/60 Hz, 10 A / 12A (refer to product label),
current
single-phase
Overvoltage category
II
Protection required
Suitable circuit breaker and a Ground Fault Circuit Interrupter (GFCI).
All protection devices must be certified according to applicable
standard. A grounded connection is essential.
Communication with other
3 USB port - 1 Ethernet port - 1 Serial port
devices
Features
Fully micro-processor controlled, process evaluation system
according to EN13060.
Programmable standby mode.
Max. heat output
3000 kJ/h
Residual-current device
30 mA or less, if required by local regulations
Automatic circuit breaker
10 A (version labelled 10 A)
(min.)
12 A (version labelled 12 A)
LISA  | Instructions for use |  VA131 ENG Rev. 05 |  31/07/2018 |  © 2018 W&H Sterilization Srl
INSTALLATION REQUIREMENTS
Working temperature
From +5 °C to +40 °C (from +41 °F to +104 °F)
Working relative humidity
Max. RH 80% up to 31 °C (88 °F), linearly decreasing to 50% at 40
°C (104 °F)
Storage temperature / rel.
From -20 °C to +60 °C (from -4 °F to +140 °F )/0–90 % (with empty
humidity
tanks)
Max altitude
3000 m asl
Min. atmospheric pressure
0.6 bar (8.7 psi)
Overall dimensions
W: 465 mm/H: 449 mm/D: 599 mm (W: 18''/H: 17.7''/D: 23.6'')
Min. space required
W: 485 mm/H: 499 mm/D: 619 mm (W: 19.1''/H: 19.6''/D: 24.4'')
Size of the door movement
W: 410 mm/H: 400 mm/D: 360 mm (W: 16.1''/H: 15.7''/D: 14.2'')
Weight empty
LISA 17: 50.2 kg (110.7 lbs)
LISA 22: 51.7 kg (114 lbs)
Max. weight (fully loaded)
LISA 17: 65.3 kg (143.9 lbs)
LISA 22: 66.8 kg (147.3 lbs)
Weight per support area
LISA 17: 38.7 kN/m
LISA 22: 39.8 kN/m
Enviroment pollution
Degree 2
STERILIZER CHAMBER
Pressure safety valve
2.6 bar (37.7 psig)
Safety thermostats
180 °C (356 °F)
Total volume
LISA 17: 17 l/Ø: 250 mm/D: 362 mm (4.5 gal, Ø: 9.8''/D: 14'')
LISA 22: 22 l/Ø: 250 mm/D: 440 mm (5.8 gal, Ø: 9.8''/D: 17'')
Technical data
2
2
115