INTERPRET THE TEST RESULT
Indicator
LISA  | Instructions for use |  VA131 ENG Rev. 05 |  31/07/2018 |  © 2018 W&H Sterilization Srl
What happened
The entire surface of the indicator sheet has changed colour.
Certain ares of the indicator sheet have not changed colour since there was an air
pocket during the cycle due to sterilizer malfunction.
Sterilizer tests
Test
What to do next
passed
Yes
-
No
Repeat the test. If it fails
repeatedly, call technical service.