Technical Data; Sterilization Cycles - W&H Lisa VA131-17 Instructions For Use Manual

Hide thumbs Also See for Lisa VA131-17:
Table of Contents

Advertisement

Technical data

CONTENTS
This section deals with the following subjects:

Sterilization cycles

Sterilization cycle phases
Technical data
Test validation points
Diagrams
Water quality
Accessories, spare parts, consumables
Authorized W&H service partners
Sterilization cycles
WARNINGS
For your safety and for the safety of your patients:
WARNING!
Never process objects different from those specified in the cycle program table and never exceed the maximum load weight limits
specified in it as this could impair the sterilization process. Such actions could result in non-sterile conditions at the end of the cycle,
could expose people to the hazard of cross-infections, are considered as an improper use of the sterilizer for which the manufacturer
cannot be hold responsible.
All indications of sterile load or successful completion of the cycle that are given on the display at the end of the cycle are not valid if
the type and quantity of the load are not complied with.
Processing bagged items with the S Fast 134 cycle will result in wet bags/pouches at the end of the cycle, exposing the items to
contamination due to improper storage. The display reminds the maximum permitted load before starting a cycle.
108
LISA  | Instructions for use |  VA131 ENG Rev. 05 |  31/07/2018 |  © 2018 W&H Sterilization Srl
108
112
115
116
117
118
119
122