Technical Data; Sterilization Cycles - W&H Lyla Instructions For Use Manual

Table of Contents

Advertisement

Technical data

CONTENTS
This section deals with the following subjects:

Sterilization cycles

Sterilization cycle phases
Technical data
Recommendations for validation
Diagrams
Water quality
Accessories, spare parts, consumables
Authorized W&H service partners
Sterilization cycles
WARNINGS
For your safety and for the safety of your patients:
WARNING!
Never process objects different from those specified in the cycle program table and never exceed the maximum load weight limits
specified in it as this could impair the sterilization process. Such actions could result in non-sterile conditions at the end of the
cycle, could expose people to the hazard of cross-infections, are considered as an improper use of the sterilizer for which the
manufacturer cannot be hold responsible.
All indications of sterile load or successful completion of the cycle that are given on the display at the end of the cycle are not valid
if the type and quantity of the load are not complied with.
Processing bagged items with the S Unwrapped 134 cycle will result in wet bags/pouches at the end of the cycle, exposing the
items to contamination due to improper storage. The display reminds the maximum permitted load before starting a cycle.
96
Lyla  | Instructions for Use |  RIN-210 ENG Rev02 |  15/09/2022 |  © W&H Sterilization Srl
96
102
105
107
108
109
110
115

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rin-210-17Rin-210-22

Table of Contents