W&H Lisa VA131-17 Instructions For Use Manual page 68

Hide thumbs Also See for Lisa VA131-17:
Table of Contents

Advertisement

Sterilization cycles
Tap
or
to increase or
5
decrease the number of label to be
printed.
Tap
to save the set number for
6
the next time.
Tap
to print the labels required.
7
SAVE ALL THE CYCLE REPORTS ON THE USB PEN DRIVE
The number of reports that can be saved on the USB pen drive
depends upon the USB capacity. To save all the cycle reports:
In the homepage tap
1
Tap
: after the confirmation all
2
sterilization cycle reports are
stored in the USB.
68
>
>
SET THE REMOTE FOLDER FOR SAVING THE REPORTS
To activate the remote storage and set the necessary parameters
do the following:
In the homepage tap
1
>
Tap
to enable the remote
2
data storage:  the first four fields in
the page and the check box turn
dark grey.
In Path enter the name of the
3
shared folder followed by the sub-
folder name, if any, where to save
reports. Do not enter the full path.
Note: The folder name must include
letters and numbers only. Do not
use other characters like space-bar,
slash, accent, etc.
Enter the host name or the IP
4
address: if the data are complete,
the fields highlight.