Maintenance
OPEN THE DOOR IN EMERGENCY
A video animation is available on https://video.wh.com/en_global/dental-faq.
Unplug the sterilizer and wait at least three hours.
1
Extract the dust filter and remove the auxiliary cable loom.
2
Plug two batteries into the connectors.
3
Firmly open the service door.
4
Holding the service door pulled, plug the plastic connector into the socket behind
5
the bacteriological filter.
As soon as the door opens, unplug the plastic connector to prevent system
6
overload and consequent damage.
Put the auxiliary cable loom back in the dust filter handle and insert the dust
7
filter in the sterilizer.
96
LISA  | Instructions for use |  VA131 ENG Rev. 05 |  31/07/2018 |  © 2018 W&H Sterilization Srl