The electrical power supply to the sterilizer must fulfill all applicable
standards in the country of use, and must comply with the data
label on the back of the sterilizer.
WATER CONNECTIONS
The water supply system must deliver demineralized or distilled
water meeting the standard requirements. Do not add any
chemical/additive to the water.
The manufacturer's warranty is void if the sterilizer was used with
water containing either chemical additives, or contaminant levels
exceeding those listed in these instructions. See "Feed water
specifications (EN 13060)" on page  118.
Notice: The maintenance of the external water filling system must
be done in exact accordance with the instructions for use given with
the relevant system.
LISA  | Instructions for use |  VA131 ENG Rev. 05 |  31/07/2018 |  © 2018 W&H Sterilization Srl
LAN CONNECTION
Insert the plug of a standard
1
Ethernet cable in the
Ethernet port.
WI-FI CONNECTION
For the Wi-Fi connection proceed as follows:
Insert the Wi-Fi key in the USB port.
1
Read the Instructions for use provided with the Wi-Fi key.
2
Getting started
Insert the other end of the
2
cable in the Ethernet wall
socket: when the sterilizer
will be switched on it will
connect automatically to the
LAN.