Technical data
Usable space
LISA 17: 11.5 l/W: 195 mm/H: 195 mm/D: 297 mm (3 gal/W:
7.7''/H: 7.7''/D: 11.7'')
LISA 22: 15 l/W: 195 mm/H: 195 mm/D: 390 mm (4 gal/W:
7.7''/H: 7.7''/D: 15.4'')
Bacteriological filter
0.3 µm
STEAM GENERATOR
Pressure safety valve
4 bar (58.01 psig)
Safety thermostats
230 °C (446 °F)
DISTILLED OR DEMINERALIZED WATER
Water quality
Fulfilling EN 13060 Ann. C (conductivity < 15μS/cm)
Average water consumption 0.32 to 0.65 litres/cycle (0.08 to 0.17 gal/cycle)
Tank volume
Clean water 4 l (1.06 gal)
Used water 3.5 l (0.92 gal)
116

Test validation points

RECOMMENDATIONS FOR VALIDATION
For further details please refer to the Qualification/Validation guide
for sterilization cycles of W&H sterilizer.
LISA  | Instructions for use |  VA131 ENG Rev. 05 |  31/07/2018 |  © 2018 W&H Sterilization Srl
Part
Description
Hottest points
Coldest points