W&H Lisa VA131-17 Instructions For Use Manual page 63

Hide thumbs Also See for Lisa VA131-17:
Table of Contents

Advertisement

STERILIZATION CYCLE PAGE
Following are the information displayed while a cycle is running:
END OF A STERILIZATION CYCLE
At the end of the S Fast 134 cycle, the sterilizer could perform two
drainage periods of about 30 seconds each and the message
"Draining" is displayed.
LISA  | Instructions for use |  VA131 ENG Rev. 05 |  31/07/2018 |  © 2018 W&H Sterilization Srl
Part
Description
A
Sterilization cycle name
B
ECO DRY mode enabled
C
Countdown clock (time until
the cycle completion)
D
: chamber pressure
: cycle counter
: chamber temperature
E
Date and time and door
securely locked symbol
F
Current cycle phase
G
Stop button
H
Button to open the cycle
information page
When a cycle is successfully finished, the "Cycle completed"
message appears on the screen. To end the cycle:
Tap
to view the cycle
1
summary or tap to view the
cycle parameters. See "View
the cycle parameters" on the
previous page.
Tap OPEN to open the door: the
2
door unlocks and the home
page appears.
Note: if an error message appears
see "Troubleshooting" on page  103
CAUTION! Hot surfaces.
Burnings. Do not touch the
chamber, the internal side
of the door and the internal
fittings. Use the tray holder
or cassette holder or
gloves for high
temperatures or adequate
protection to remove the
load!
Open the chamber door.
3
Remove the load and stock it.
4
Sterilization cycles