W&H Lisa VA131-17 Instructions For Use Manual page 46

Hide thumbs Also See for Lisa VA131-17:
Table of Contents

Advertisement

Sterilizer tests
In the homepage tap
.
3
Tap
and enter your credentials if required: the chamber door locks.
4
Wait until the end of the test and tap OPEN: the chamber door unlocks.
5
Enter your credentials if required.
6
Open the chamber door, extract the tray using the tray holder and take the test
7
pack. The test pack can be wet outside.
CAUTION! Risk of burn. The test pack is very hot at the end of the cycle. Wear
appropriate PPE (e.g. gloves).
Remove the indicator sheet from the center of the test pack and check the
8
change in colour. See "Interpret the test result" on the next page
46
LISA  | Instructions for use |  VA131 ENG Rev. 05 |  31/07/2018 |  © 2018 W&H Sterilization Srl

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lisa va131-22

Table of Contents