Technical data
Minimum duration of the drying phase (set by the user)
Load type
TOTAL CYCLE DURATION
The total cycle time includes the drying time and may vary according to different elements, such as the following:
type of load (solid or porous)
n
load weight
n
the duration of the drying phase (if the ECO DRY mode is disabled)
n
other factors
n
Values and cycle names could be different depending on country requirements.
B Universal 134
B Extended 134
110
All unwrapped, bagged, single/double wrapped items:
n
n
n
n
Empty
LISA 17
LISA 22
21'
21'
36'
36'
Sterilization cycles
B Universal 134
B Extended 134
16'
16'
Solid
Hollow A (Narrow lumen)
Hollow B (simple hollow item)
Porous
Load
Full
LISA 17
LISA 22
44'
49'
60'
64'
LISA  | Instructions for use |  VA131 ENG Rev. 05 |  31/07/2018 |  © 2018 W&H Sterilization Srl
B Universal 121
S Fast 134
24'
-
Unwrapped items:
Solid
n
Hollow B
n
(simple
hollow
item)
Dental load
n
(turbines
and
handpieces
family, ...)
Typical
LISA 17
LISA 22
30'
31'
46'
47'