Diagnostic
Code
2xx
Switch the sterilizer OFF and ON.
Wait for the chamber to cool down. Reset the thermal
overload (see "Extraordinary maintenance" on page  91).
3xx
Check the door seal. Clean or replace it if necessary.
Clean the chamber face side.
Clean the chamber filter.
Check if chamber filter is properly locked in the cap.
Check the load does not exceed the MAXIMUM WEIGHT
LIMITS.
4xx
Clean water error (bad quality or low water level). Drain
and refill the clean water tank.
5xx
Check if there are hurdles on the door locking area
(chamber rack, loads, objects, ...).
Check the door seal (wrong placed).
Check if the door can move freely without touching the
trays or the load when closing.
Switch the sterilizer OFF and ON.
990
The cycle has been aborted by the user.
100
Description and actions
Repeat the cycle.
If the problem persists, call service.
Repeat the cycle.
If the problem persists, call service.
Repeat the cycle.
If the problem persists, call service.
Repeat the cycle.
If the problem persists, call service.
Re-process the load.