Sterilizer tests
In the homepage tap
.
3
Tap
and enter your credentials if required: the chamber door locks.
4
Wait until the end of the test and tap OPEN: the chamber door unlocks.
5
Enter your credentials if required: a message informs if the test passed or failed.
6
If the test failed, see "What to do when the test failed" below
WHAT TO DO WHEN THE TEST FAILED
Check, clean or replace the door gasket.
1
Clean the chamber face side and the chamber filter.
2
Repeat the Vacuum test. See "Carry out the test" on the previous page.
3
If the test fails repeatedly, call technical service.
4