Sterilizer tests
Open the chamber door, extract the tray using the tray holder and take the tube.
9
CAUTION! Risk of burn. The test pack is very hot at the end of the cycle. Wear
appropriate PPE (e.g. gloves).
Unscrew the tube capsule and remove the indicator strip.
10
Check the change in colour. See "Interpret the test result" below
11
INTERPRET THE TEST RESULT
Indicator
WHAT TO DO NEXT
Compile the Helix test documentation form to trace the effectiveness of the sterilization cycle during the whole lifespan of your sterilizer. See
"Helix test documentation form" on page  128.
50
What happened
The indicator strip has turned dark. The air was completely removed from
the capsule.
Part of the chemical indicator strip has not turned dark. The air removal
from the capsule was not complete.
Test
passed
Yes
No
LISA  | Instructions for use |  VA131 ENG Rev. 05 |  31/07/2018 |  © 2018 W&H Sterilization Srl
What to do next
-
Repeat the test. If it fails repeatedly,
call technical service.