Sterilization cycles
PROGRAM A SEQUENCE OF CYCLES TO RUN AUTOMATICALLY
It is possible to program the sterilizer in such a way that a
sequence of cycles runs automatically at a preset date and time
(e.g. if you want to run a vacuum test plus a HELIX test in the early
morning before office hours). Maximum two cycles can be
programmed.
To program the sterilizer proceed as follows:
Load the chamber and close the
1
chamber door.
In the homepage tap
2
>
Tap
to add a cycle to the list
3
above.
60
>
>
Tap
and then the cycle you want
4
to add.
Tap
: the cycle is automatically
5
listed in the previous page.
Tap the date/time to set the start of
6
the cycle.
LISA  | Instructions for use |  VA131 ENG Rev. 05 |  31/07/2018 |  © 2018 W&H Sterilization Srl