W&H Lisa VA131-17 Instructions For Use Manual page 102

Hide thumbs Also See for Lisa VA131-17:
Table of Contents

Advertisement

Diagnostic
Message/Alert
Possible leak detected.
Please run Vacuum Test.
Remote data storage
Please check settings.
Remote data storage
File lost.
Door Gasket replacement is due.
Have you replaced the Door
Gasket?
HEPA filter replacement is due..
Have you replaced the HEPA
filter?
Dust Filter replacement is due.
Have you replaced the Dust
Filter?
LISA has completed 4000 cycles.
Please call your technical support
to organize service.
102
Description
Air was detected in the chamber: a vacuum leak is
suspected. The cycle was completed but a vacuum
test is required.
Error when saving the files in the network.
Some cycle reports couldn't be saved in the
network because the sterilizer memory is limited
(see "Sterilization cycle report" on page  65).
These messages advise that one consumable
must be replaced.
This message advises you that the target of 4000
cycles has been achieved and the relevant
maintenance step is to be performed.
Run a vacuum test.
Call for service if an anomaly is detected.
Check the settings, the network resource and the memory space available on it.
No action possible. The cycles were overwritten in the sterilizer database before being
saved, and there is no way to recover them.
Replace the consumable and tap
page  73).
If you don't replace the consumable, press
In this case, you may still use the sterilizer but the message will appear again after
some cycles.
CAUTION! Operating the sterilizer with expired consumables could be
dangerous and could damage the sterilizer.
Call technical service for the 4000 cycle maintenance.
LISA  | Instructions for use |  VA131 ENG Rev. 05 |  31/07/2018 |  © 2018 W&H Sterilization Srl
Action
to reset the counter (See "Maintenance" on
.

Hide quick links: