Conductivity measurement

Measurement

Set the cell constant C, the temperature function TC
and the reference temperature correctly first !
The conductivity menu comprises the following points:
1.
Adjust ("C") or determine ("C Cal") cell constant with
subsequent selection of meas. range. Only the meas. ranges
which are valid for the actual cell constant are displayed.
2.
Adjust temperature function ("TC") or determine ("TC Cal").
3.
Set reference temperature Tref.
4.
Documentation.
BA61101/08.00/IWa/MLP5-E2
Âmenu
25.4°C
Á®
Ãmeth.
nlF 25
275,9
Âmenu
® µS/cm 25.4°C
nlF 25
Âmenu
Á®
25.4°C
Ãmeth.
nlF 25
Rinse meas. cell with a part
of sample and immerse into
the solution.
Menu conductivity
measurement.
Conductivity measurement
runs continuously.
Back to menu
conductivity
measurement.