BA61101/08.00/IWa/MLP5-E2
Ammonium
ISE CAL NH¿-N
È
C1= 10.00m¥/L
Ã
25.4°C
ISE CAL NH¿-N
Ï
C1=009. 0m¥/L
Ã
25.4°C
Rinse electrode and temperature probe and immerse
into first standard solution.
ISE CAL NH¿-N
C1= 9.70 m¥/L
U1= -
80.4mV
ÃC2
25.4°C
ISE CAL NH¿-N
É
C2= 00.00m¥/L
Ã
25.4°C
Rinse electrode and temperature probe and immerse
into second standard solution.
ISE CAL NH¿-N
C2= 95.00m¥/L
U2= - 139.1mV
ÃCalDat
25.4°C
Select calibration.
Enter concentration of the
first standard solution.
: Select figure
: Set figure.
Start measurement. Wait
for stable test values.
Enter concentration of the
second standard solution.
: Select figure
: Set figure.
Start measurement. Wait
for stable test values.
79