Reset

Reset to to default settings is possible in single steps. "Reset all" is
the sum of all single steps.
Own methods and test values remain stored.
Below all possibilities are listed. Ã starts the selected Reset.
+
BA61101/08.00/IWA/MLP5-E2
Configuration
ÂMenü
ÃReset
ÂMessen
Reset all
¡Photo/Foto
pH
ÂMessen
Photo/Foto
¡pH
®
Switch off instrument.
Switch on instrument and
simultaneously depress à .
After acoustic signal release
à .
¡ Reset Photometry.
¡ Reset pH/mV/ISE-
measurement.