Troubleshooting
Measurement with blank value adjustment
BA61101/08.00/Pla/MLP5-E2
Measurement:
Measured value exceeds measuring
range of the reagent set.
The displayed measured value (e.g. 11.07
mg/L) is an assessed value.
Action:
- Give result as assessed value
- Use reagent set with higher measuring
range
- Select vial/cuvette with thinner layer.
Measurement:
Measured value exceeds measuring
range of the reagent set.
It is not possible to compute an assessed
value.
Action:
- Pre-dilute sample
- Use reagent set with higher measuring
range
Sample blank value adjustment:
Small influence of the sample blank value.
The sample blank value adjustment ist
only necessary for very precise
measurements or for measurements with
very low concentrations.
215