Performance of instrument zero adjustment

Perform zero adjustment of the instrument with
mains power supply only!
+
208
INSTRUMENT ZERO ADJUSTMENT ONLY AFTER
CHANGING LAMP or FILTER, or AFTER
RECEIVING ERROR MESSAGE "ERROR 1"!
In delivery condition all zero adjustments of the
instrument have been carried out carefully and stored.
For MEASUREMENT in practice NO ZERO
ADJUSTMENT required!
The values of the zero adjustments remain stored
until the next adjustment.
The instrument doesn't recognize a wrong zero
adjustment ! Zero adjustment should be carried out
by skilled persons only.
Maintenance
Connect line adaptor (see
chapter mains power
supply).
Switch on instrument and
simultaneously press
Release À
Lamp : Zero adjust after
À
changing the lamp.
l = 445 nm, filter pos. 1.