Ammonium Measurement; Calibration Of The Ammonium Electrode - wtw MultiLab P5 Operation Manual

Multiparameter instrument for the total water and wastewater analysis
Table of Contents

Advertisement

Calibration of the ammonium electrode*

*
e. g. WTW electrode NH500/2
Basic adjustment without calibration.
Two-point calibration without temperature compensation.
From two pairs of values the instrument calculates the axial intercept
and the slope of the characteristics.
Basis of the calculation is the Nernst equation U=U0+Slg(c).
The temperature difference between calibration and sample solution
should not exceed 2 K. The displayed temperature only has a control
function.