Wait
The zero adjustment after changing the lamp of the MultiLab P 5
must be carried out separately for all used vial/cuvette types and
filter positions.
Attend to the below mentioned min. filling capacity for each
vial/cuvette type for an orderly zero adjustment !
Cuvette
l 14 mm vial
¨ 10 mm rectangular cuvette
¨ 20 mm rectangular cuvette
¨ 50 mm rectangular cuvette
BA61101/08.00/Pla/MLP5-E2
Maintenance
Insert vial/cuvette containing distilled water (clean,
not scratched!).
Attend to minimum filling capacity (s.b.)!
Start zero adjustment .
Zero adjustment is finished.
Min.
filling capacity
[ml]
2.6 ml
2.0 ml
4.0 ml
10 ml
Min.
filling height
[mm]
20 mm
20 mm
20 mm
20 mm
209