wtw MultiLab P5 Operation Manual page 209

Multiparameter instrument for the total water and wastewater analysis
Table of Contents

Advertisement

Wait
The zero adjustment after changing the lamp of the MultiLab P 5
must be carried out separately for all used vial/cuvette types and
filter positions.
Attend to the below mentioned min. filling capacity for each
vial/cuvette type for an orderly zero adjustment !
Cuvette
l 14 mm vial
¨ 10 mm rectangular cuvette
¨ 20 mm rectangular cuvette
¨ 50 mm rectangular cuvette
BA61101/08.00/Pla/MLP5-E2
Maintenance
Insert vial/cuvette containing distilled water (clean,
not scratched!).
Attend to minimum filling capacity (s.b.)!
Start zero adjustment .
Zero adjustment is finished.
Min.
filling capacity
[ml]
2.6 ml
2.0 ml
4.0 ml
10 ml
Min.
filling height
[mm]
20 mm
20 mm
20 mm
20 mm
209