O 2 Reference: Calibration Against A Known (Titrated) Concentration Value - wtw MultiLab P5 Operation Manual

Multiparameter instrument for the total water and wastewater analysis
Table of Contents

Advertisement

Oxygen measurement
O
Reference: Calibration against a known (titrated)
2
concentration value
Information on the solubility function:
The solubility of oxygen depends on the solvents, the dissolved
agents and the temperature. Oxygen partial pressure and
temperature are measured. The concentration is calculated according
to the solubility function for pure water (or seawater respectively)
stored in the instrument.
The solubility factor indicates the relative concentration difference to
pure water (or seawater).
Except for pure water the displayed concentration value is valid for
solutions of the same composition at calibration temperature only.
BA61101/08.00/IWa/MLP5-E2
Immerse oxygen probe into solution with known
oxygen concentration!
Care for incident flow!
ÂOº
Sal
¡OxiCal
HºO-Cal
ÂOº
Oº= 10.2m¥/L
É
F=
.000
ÂOº
Oº=
9.7m¥/L
È
F=
.954
Select calibration.
With
or
select
calibration procedure O
reference.
Confirm selection.
Display of the current
oxygen concentration.
Change factor until the
correct oxygen
concentration is set.
: Select figure
: Set figure.
2
115

Advertisement

Table of Contents
loading

Table of Contents