Changing the lamp

BA61101/08.00/IWA/MLP5-E2

Maintenance

Switch off the instrument!
A zero adjustment must be
carried out after changing the
lamp!
Turn out screw.
Remove lid.
Turn out screw.
Remove lamp.
Insert new lamp.
Do not touch glass bulbs of the
lamp with your fingers!
Screw on lamp.
Insert lid.
Screw on lid.