wtw MultiLab P5 Operation Manual page 92

Multiparameter instrument for the total water and wastewater analysis
Table of Contents

Advertisement

92
Ionselective measurement
ISE CAL
Ê
C1= 11. 0
Ã
25.4°C
Rinse electrode and temperature probe and immerse
into first standard solution.
ISE CAL
C1=
11.20
U1=
10.8mV
ÃC2
25.4°C
ISE CAL
C2=
É
00.00
Ã
25.4°C
Rinse electrode and immerse into the second
standard solution.
ISE CAL
C2=
102.10
U2=
-45.3mV
ÃCalDat
25.4°C
ÂISE
ISE CAL
S=
-57.5mV/D
U0=
-102.3mV
ÂISE
ISE CAL
C1=
11.20
C2=
102.10
Enter the concentration of
the first calibration standard
solution.
: Select figure
: Set figure.
Start measurement. Wait
for stable test values.
Enter the concentration of
the second calibration
standard solution.
: Select figure
: Set figure.
Start measurement. Wait
for stable test values.
Display of calibration data.
Indication of used standard
solutions.
switch to-and-fro between
the calibration data.
BA61101/08.00/IWa/MLP5-E2