Troubleshooting

Accu
Control elements
BA61101/08.00/Pla/MLP5-E2
Symbol for state of charge Æ is lit:
Charge accumulator.
Charging the accu: Æ is faded out
approx 3 min. after connection of
mains plug transformer.
The symbol for state of charge Æ
replaces the marking sign ¡ in the
selected menus.
Display is dark.
Instrument should function again after
connection of the mains plug transformer.
If not send instrument to WTW for repair.
Wrong positioning of the cuvette:
The rectangular cuvette is not inserted
right-flushed in the cuvette compartment.
Wrong cuvette:
Rectangular cuvette instead of vial has
been used for reaction vial test.
213