Ionselective measurement
Calibration
Two-point calibration without temperature compensation.
From two pairs of values the instrument calculates the axial intercept
and the slope of the characteristic.
The temperature difference between calibration and sample solution
should not exceed 2 K.
BA61101/08.00/IWa/MLP5-E2
Âmenu
ÀCal
25.4°C
ÁpH
Ãmeth.
AR TP
Âmeas.
14697 Äa-Ten
¡pH
mV
Âmeas.
ISE F
¡ISE
14779 ÅHS
Âmenu
ÀCal
25.4°C
ÁISE
Ãmeth.
ISE CAL
C1= 0 0.00
É
Ã
25.4°C
Position 10: pH menu.
Method change.
ISE
= Select ionselective
measurement.
Entry to ISE menu.
TP
Select calibration.
91