The recognition of an error leads to error message "INVALID
COMMAND".
The following inputs lead to error message "INVALID COMMAND" as
well:
- Slope = 0 or > 32000
- Zero pointt > 32000
- Beginning of meas. range < 0 or > 32000
- End of meas. range < 0 or > 32000
- End of meas. range £ Beginning of meas. range
- Ref. cuvette different 10 mm, 14 mm, 20 mm or 50 mm
- Wave length does not correspond to the instrument
- Resolution not 0.1, 0.01, ...
After successful input the instrument displays the received method :
Totally 50 own methods are possible, max. 40 per filter position.
The 51st own method on one filter causes "Meth. Error"; in this
case another method should be erased before input of the new
method.
BA61101/08.00/IWa/MLP5-E2
Documentation
Âmenu
mmol/L
ÁTest1
Ãmeth.
The instrument limits independently the meas.
range (no remote indication!), when the entered
method data allow extinction values > 3 resp. > 2,5
(at 340 nm).
NH¿-N
177