216
Troubleshooting
Sample blank value adjustment:
Very big influence of the sample blank
value. The measuring accuracy is
considerably reduced.
Cause/Action:
- Reduce sample blank value
- Sample turbid / Filter sample,
e. g. ultrafiltration
- Sample colored / Remove coloring,
e. g. oxidation
Sample blank value adjustment:
Very big influence of the sample blank
value. Measurement is impossible.
Cause/Action:
- Reduce sample blank value
- Sample turbid / Filter sample,
e. g. ultrafiltration
- Sample colored / Remove coloring,
e. g. oxidation
Sample blank value adjustment:
Sample blank value reduces the
measuring range for the measurement.
Action:
- Decrease sample blank value
- Dilute sample
BA61101/08.00/Pla/MLP5-E2