Dissolved Oxygen - wtw MultiLab P5 Operation Manual

Multiparameter instrument for the total water and wastewater analysis
Table of Contents

Advertisement

Dissolved oxygen O
Parameters
Meas. principle
Meas. ranges (resolution)
Accuracy
Temperature
compensation
only slow NTC with OxiCal
Air pressure correction
AutoBar
Salinity correction
Temperature
Meas. range
Resolution
Accuracy temperature
(at 15°C..35°C ambient
temperature)
Sensor exchange error
Temperature range
®
TriOxmatic
300
Functions
Calibration DIN EN 25814
BA61101/08.00/IWa/MLP5-E2
Technical data
mg/L O
temperature continuously
potentiostatic 3-electrode system
0.0 to 300.0 mg/L (0.1 mg/l)
0%..999 % (1%) saturation.
In the range 0.0 mg/L..30.0 mg/L the
resolution is selectable between 0.1
mg/L and 0.01 mg/L.
In the range 0.0%..100.0% the
resolution is selectable between 0.1%
and 1%.
0.5 % of meas. value
(0.01 mg/L resolution)
automatically with double thermistor
system
automatically from 850 to 1080 mbar
0..99, manually adjustable
-5.0°C..60°C
0.1K
± 0.1K ± 1 digit
<0.2K at -5.0°C .. 60.0 ° C
<0.4K at 60.0°C .. 110.0 ° C
0°C..50°C
AutoReg, membrane control, AutoBar,
AutoRead, OxiCal, H
BSB5/BOD
OxiCal with OxiCal P or OxiCal SL
H
(T
, T
, p)
2
1
2
, mbar pO
2
O-Cal with air-saturated water.
2
, % saturation,
2
O-Cal,
2
241