86
ISE CAL
Ñ
C1=009.9 m¥/L
Ã
Rinse electrode and temperature probe and immerse
into first standard solution.
ISE CAL
C1= 9.99 m¥/L
U1=
10.8mV
ÃC2
25.4°C
ISE CAL
É
C2= 00.00m¥/L
Ã
Rinse electrode and temperature probe and immerse
into second standard solution.
ISE CAL
C2=102.11m¥/L
U2=
-45.3mV
ÃCalDat
25.4°C
ÂISE F
ISE CAL
S=
-57.5mV/D
U0=
-102.3mV
Fluoride
Enter the concentration of
F-
the first standard solution.
: Select figure
: Set figure.
F-
Start measurement. Wait
for stable test values.
Select concentration of the
F-
second calibration standard
solution.
: Select figure
: Set figure.
Start measurement. Wait
F-
for stable test values.
Display of calibration data.
F-
BA61101/08.00/IWa/MLP5-E2