J-Lite JLA-16.250 Rotor
REMOVAL AND SAMPLE RECOVERY
14
!
CAUTION
Make sure that the rotor is properly seated on
and securely fastened to the drive hub before
each run.
When the rotor is properly loaded and installed in the centrifuge,
you are ready to perform the run. Refer to the applicable centrifuge
instruction manual for operating instructions.
NOTE
Line voltage fluctuations can cause variations in
acceleration and deceleration times.
!
CAUTION
If disassembly reveals evidence of leakage, you
should assume that some fluid escaped the rotor.
Apply appropriate decontamination procedures
to the centrifuge and accessories, as required.
Dual-Locking Lid
1. Place one hand on the rotor to keep it from turning. Grasp the
tie-down knob with your other hand and turn it to the left (counter-
clockwise), while pulling up, until the stem disengages from the
drive hub.
2. Place your hands on the rotor and lift the rotor straight up off the
drive hub. Do not lift the rotor by the tie-down knob.
3. Place the rotor on a counter or bench top under a safety hood
(if required).