J-Lite JLA-16.250 Rotor
16
container.) To minimize the possibility of leakage from capped tubes or
bottles, load tubes or bottles with sample, secure the caps, and precool
the loaded, sealed labware to run temperature before beginning the run.
Refer to Appendix A in Rotors and Tubes for information on the chem-
ical resistances of tube and accessory materials.
Temperature Limits
• Plastic containers have been tested for use at temperatures between 2
and 25°C. For centrifugation at other temperatures, pretest
containers under anticipated run conditions.
• If plastic containers are frozen before use, make sure that they are
thawed to at least 2°C before centrifugation.