TEMPERATURE
To ensure that the JLA-16.250 rotor reaches the required temperature
during the run, follow the appropriate instructions below for the
J centrifuge model being used.
Avanti J Series Centrifuges (Except Avanti J-E)
Enter the run temperature according to the instructions in your centri-
fuge instruction manual. No additional input is required.
For Avanti J-E Series Centrifuge
Use the following temperature offset table to set the desired sample
temperature.
Table 1. Temperature Settings for Model J-E High Performance
Centrifuge. At speeds above 13,000 rpm, use the indicated temperature
setting for the desired sample temperature. Interpolate if intermediate values
are required. At some speed and temperature combinations, the centrifuge
may be unable to achieve the desired sample temperature.
Rotor
Speed
(rpm)
–10
16 000
N
14 000
N
13 000
–10
"N" indicates that the rotor cannot achieve the required temperature at this speed.
Required Sample Temperature (°C)
–4
2
4
6
N
N
N
N
–6
0
2
6
–4
2
4
6
J-Lite JLA-16.250 Rotor
8
10
20
6
8
20
8
10
20
8
10
20
11