Beckman Coulter J-Lite JLA 16.250 Instruction Manual page 20

Fixed angle rotor used in beckman coulter j2 and avanti j series centrifuges
Table of Contents

Advertisement

J-Lite JLA-16.250 Rotor
40 000
35 000
30 000
25 000
20 000
15 000
10 000
5 000
0
0
20
Table 5. Relative Centrifugal Fields for the JLA-16.250 Rotor.
Entries in this table are calculated from the formula
RCF = 1.12r (RPM/1000)
and are then rounded to three significant digits.
Rotor
at r
Speed
max
(rpm)
(134 mm)
16 000
38 400
15 000
33 800
14 000
29 400
13 000
25 400
12 000
21 600
11 000
18 200
10 000
15 000
9 000
12 200
8 000
9 610
7 000
7 350
6 000
5 400
5 000
3 750
4 000
2 400
3 000
1 350
2 000
600
1 000
150
500
Relative Centrifugal Fields for the JLA-16.250 Rotor
2 000
4 000
6 000
2
Relative Centrifugal Field (× g)
at r
av
(90 mm)
25 800
22 700
19 800
17 000
14 500
12 200
10 100
8 160
6 450
4 940
3 630
2 520
1 610
907
403
100
38
25
8 000
10 000
12 000
Speed (rpm)
at r
min
(46 mm)
13 200
11 600
10 100
8 710
7 420
6 230
5 150
4 170
3 300
2 520
1 860
1 290
824
464
206
52
13
r max
r av
r min
14 000
16 000

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

J-lite jla-16.250

Table of Contents