J-Lite JLA-16.250 Rotor
BEFORE USING THE JLA-l6.250 ROTOR IN A MICROPROCESSOR-CONTROLLED
J2 SERIES CENTRIFUGE
8
Figure 2. Typical Examples of Arranging Bottles or Tubes in the Rotor.
Two, three, four, or six containers can be centrifuged per run,
2if they are arranged in the rotor as shown.
• If fluid containment is required, use capped bottles or tubes. It is
strongly recommended that all containers carrying physiological
fluids be capped, and not overfilled, to prevent leakage.
• If you are using an Avanti J series centrifuge, select the
JLA-16.250 rotor.
• If you are using a microprocessor-controlled J2 series centrifuge,
enter rotor code 14. Refer to your centrifuge instruction manual
for additional information. See
ROTOR IN A MICROPROCESSOR-CONTROLLED J2 SERIES
, below, if you are using the rotor for the first time.
CENTRIFUGE
If you have a microprocessor-controlled Beckman Coulter J2 series
centrifuge (model J2-MI, J2-21M, J2-21M/E, or J2-MC), you will
need to enter the JLA-16.250 rotor identification code (14) and
maximum speed (l4 000 rpm) into the instrument memory before the
first use of the JLA-16.250 rotor in that instrument. Use the following
instructions.
1. On the centrifuge panel, press and hold the [
holding [
], press [
ROTOR
display area will show 1 – – . – . This indicates that you are ready
to program the first entry, which is the rotor identification code 14.
Enter 14, again press and hold the [
and then release both keys.
BEFORE USING THE JLA-16.250
ROTOR
]. Release both keys. The rotor
PROG SAVE
] key, press [
ROTOR
] key. While
],
PROG SAVE