J-Lite JLA-16.250 Rotor
12
J2 Series Centrifuges
Enter the required run temperature and the appropriate temperature
compensation units (see Tables 1 and 2) on the centrifuge control
panel following the instructions below.
Table 2. Temperature Compensation Settings for Analog
and Microprocessor-Controlled Centrifuges (Models J2-HS,
J2-21, J2-MI, J2-21M, J2-MC, and J2-21M/E).
Interpolate if intermediate values are required.
Rotor
Speed
(rpm)
–10
14 000
N
10 000
–4
5 000
0
"N" indicates that the rotor cannot achieve the required temperature at this speed.
Table 3. Temperature Compensation Settings for the Model J2-HC
Analog Centrifuge. Interpolate if intermediate values are required.
Rotor
Speed
(rpm)
–10
14 000
N
10 000
–4
5 000
0
"N" indicates that the rotor cannot achieve the required temperature at this speed.
Analog J2 Series Centrifuges (Models J2-HS, J2-21, and J2-HC)
1. Turn the SET knob on the centrifuge control panel to the required
sample temperature.
2. Find the compensation value in Table 1 or Table 2 that corre-
sponds to the required temperature and run speed. Set the COMP
dial to that setting. (Interpolate if intermediate values are
required.)
Required Sample Temperature (°C)
2
5
10
–7
–5
–4
–3
–3
–2
0
0
0
Required Sample Temperature (°C)
2
5
10
–7
–7
–7
–3
–3
–2
0
–3
–3
20
40
–3
–3
–1
N
0
N
20
40
–5
–4
–1
N
–3
N