Automatic Storage At Intervals - wtw Multi 3430 Operating Manual

Digital ph/dissolved oxygen/conductivity meter for 3 digital ids sensors
Hide thumbs Also See for Multi 3430:
Table of Contents

Advertisement

Multi 3430
Configuring the
automatic storage
function
Settings
ba75865e02
12/2009

4.9.2 Automatic storage at intervals

The storage interval (Interval) determines the time interval between
automatic storage processes. Each storage process transmits the cur-
rent dataset to the interface at the same time.
1
Press the <STO
The menu for automatic storage appears.
Automatic data storage
ID number
Interval
Duration
Continue
0d03h00min
0
22.09.2009 08:00
You can configure the automatic storage function with the following set-
tings:
Menu item
Possible
setting
ID number
1 ... 10000
Interval
1 s, 5 s, 10 s,
30 s, 1 min,
5 min,
10 min,
15 min,
30 min,
60 min
Duration
1 min ... x min
_
> key.
1
30 s
180 min
1d17h33min
Description
ID number for the dataset series.
Storage interval.
The lower limit of the storage
interval can be restricted by the
number of free storage locations.
The upper limit is restricted by the
storage duration.
Storage duration.
Specifies after which time the
automatic storage should be ter-
minated.
The lower limit of the storage
duration is restricted by the stor-
age interval.
The upper limit is restricted by the
number of free storage locations.
Operation
Adjusted entire storage
duration
Max. available storage
duration
Graphical display of the
storage usage
75