The page up and page down keys allow you to move the display up
or down a page at a time.

Left/Right

The < CTRL> -F and < CTRL> -G keys allow you to move the
display left or right, respectively. These keys are used if the
emulator contains an external analyzer and you have an 80 column
display (The default trace display includes the external bits
information in this case, but this information cannot be displayed
on an 80 column display.)
The < CTRL> -F and < CTRL> -G keys are also used when the
width of the address or mnemonic/absolute columns is increased so
that not all the trace display data can be displayed across the screen.
The "display trace depth" command allows you to specify the
number of states that are displayed. By reducing the trace depth,
you can shorten the time it takes for the Softkey Interface to
upload the trace information. You can increase the trace depth to
view more states of the current trace.
The maximum number of trace states is 1024 when counting is
turned off, 512 when counting is on. When you initially enter the
Softkey Interface, the trace depth is 256. The minimum trace
depth is 9.
If you wish to reduce the number of states that are displayed, the
"display trace depth" command must be entered before the trace is
displayed. You cannot use this command to reduce the number of
states displayed in the current trace.