SE-601 Series Electrocardiograph User Manual
When Paper Marker is set to No,
1) The device does not roll back the recorder paper to the start point of each page at the
beginning of the printing course, and it does not advance the paper to the next black
marker at the end of the printing course.
2) Before printing, you can press the Space key to advance the paper for about 2cm.

9.3.3 Selecting Patient Information

For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Patient Info item in the
Record Setup window, and then press F1 or F2 to select Off or On.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Patient Info item in the
Record Setup window, and then press F3 or F4 to select Off or On.
In the AUTO mode, when Patient Info is set to On, the patient information will be printed in the
ECG reports.
In the AUTO mode, when Patient Info is set to Off, the patient information will not be printed in
the ECG reports.

9.3.4 Setting Speed

For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Speed item in the Record
Setup window, and then press F1 or F2 to select an option.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Speed item in the
Record Setup window, and then press F3 or F4 to select an option.
In the MANU mode, you can set Speed to 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s or
50mm/s. In the AUTO and RHYT modes, only 25mm/s and 50mm/s are available. In the R-R
mode, only 25mm/s is available.

9.3.5 Setting Gain

For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Gain item in the Record
Setup window, and then press F1 or F2 to select an option.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Gain item in the
Record Setup window, and then press F3 or F4 to select an option.
You can set the indicated height of 1mV ECG on the paper.
You can set Gain to 10mm/mV, 20mm/mV, 10/5mm/mV, AGC, 2.5mm/mV or 5mm/mV.
AGC means auto gain control. When ECG signals vary greatly, AGC can be selected to adjust
the gain automatically according to actual signals.
10/5mm/mV means that the gain of limb leads is set to 10mm/mV, while the gain of chest leads
is set to 5mm/mV.
- 83 -