SE-601 Series Electrocardiograph User Manual
System Setup
9.1.6 Specifying Recording Sequence
For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Record Sequence item in the
Work Mode Setup window, and then press F1 or F2 to select Sequential or Simultaneous.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Record Sequence item
in the Work Mode Setup window, and then press F3 or F4 to select Sequential or Simultaneous.
When Record Sequence is set to Sequential, the lead group is printed one by one in a certain
sequence. The start time of a lead group is just the end time of the previous lead group, as the
following figure shows.
When Record Sequence is set to Simultaneous, all leads are printed simultaneously. The start
time of each group is the same. All the start time is 0s, as the following figure shows.
9.1.7 Inputting Sampling Time
Press Tab or Shift + Tab to move the cursor to the Sample Time textbox in the Work Mode
Setup window, and then enter the sampling time. The range of the sampling time is 10-24s.
The sampling time is only valid in the auto style of 6×2 adjust.