EDAN SE-601A User Manual page 60

Se-601 series. electrocardiograph
Hide thumbs Also See for SE-601A:
Table of Contents

Advertisement

SE-601 Series Electrocardiograph User Manual
25mm/s (Paper Speed)
10mm/mV (Gain)
60 (Heart Rate)
V1.0 (Version number)
NOTE: The version number in the ECG reports is just for reference. The real version
number can be found on the screen of the electrocardiograph when you turn on
the device.
Figure (b) shows: Patient Information, Measure Information, Minnesota Code, Diagnosis
Information, Average Template, Report Reviewed by, Doctor Name, Current Date and Current
Time, 0.15Hz (DFT Filter), 100Hz (Lowpass Filter), AC50 (50Hz AC Filter), 25mm/s (Paper
Speed), 10mm/mV (Gain), V1.0 (Version number).
Patient Information includes:
ID, Name, Age, Gender, Height, Weight, BP, Race, Medication, Ward NO, Tel (Extra
Information)
Measure Information includes:
HR (Heart Rate)
P Dur----P wave duration: the average P-wave duration from several selected dominant beats;
PR int----P-R interval: the average P-R interval from several selected dominant beats;
QRS Dur----QRS complex duration: the average QRS complex duration from several selected
dominant beats;
QT/QTC int----Q-T interval: the average Q-T interval from several selected dominant beats
Normalized QT interval;
P/QRS/T axis----Dominant direction of the average integrated ECG vectors;
RV5/SV1 amp----The maximum value of the amplitude of R and R' wave from the average beat
of lead V5 / The maximum absolute value of the amplitude of S and S' wave
from the average beat of lead V1;
RV5+SV1 amp----Sum of RV5 and SV1;
RV6/SV2 amp----The maximum value of the amplitude of R and R' wave from the average beat
of lead V6 / The maximum absolute value of the amplitude of S and S' wave
from the average beat of lead V2.
Diagnosis Information:
Diagnosis information shows the auto diagnosis result.
Average Template:
Average template shows the average value of 10s sampled ECG signals of every lead.
The broken lines on the template are position markers. They respectively mark the start and end
points of the P wave and the QRS wave, and the end point of the T wave.
- 53 -