SE-601 Series Electrocardiograph User Manual
A1.5 Performance Specifications
Recording
Recorder:
Printing Density
Recorder Paper:
Effective Width:
Paper Speed:
Accuracy of data:
HR Recognition
Technique:
HR Range:
Accuracy:
ECG Unit
Leads:
Acquisition Mode:
A/D Resolution:
Time Constant:
Frequency Response:
Gain:
Input Impedance:
Input Circuit Current:
Input Voltage Range
Calibration Voltage:
DC Offset Voltage:
Noise:
Thermal dot-matrix recorder
8 dots per mm / 200 dots per inch (amplitude axes)
40 dots per mm / 1000 dots per inch (time axes, @ 25
mm/s)
Folded thermal paper: 110mm×140mm×144pages
104mm
5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s
(±3%)
±5% (x-axis), ±5%(y-axis)
Peak-peak detection
30 BPM ~300 BPM
±1 BPM
12 standard leads
simultaneously 12 leads
24 bits
≥3.2s
0.05Hz ~ 150Hz (-3dB)
2.5mm/mV, 5mm/mV, 10mm/mV, 20mm/mV, 10/5mm/mV,
AGC
≥50MΩ (10Hz)
≤0.01μA
≤±5mVpp
1mV±2%
±600mV
≤12.5µVp-p
- 117 -
Technical Specifications