4.5.2 Disposable Electrodes

Disposable Electrode
Disposable electrodes must be used together with alligator clips.
The electrodes' positions on the body surface are shown in the following table and figures.
American
European
Label
Label
RA
R
LA
L
RL
N or RF
LL
F
V1
C1
V2
C2
V3
C3
V4
C4
V5
C5
V6
C6
Electrode Placement
Right deltoid
Left deltoid
Above the right ankle (Alternate placement, upper leg as close to
torso as possible)
Above the left ankle (Alternate placement, upper leg as close to torso
as possible)
Fourth intercostal space at the right border of the sternum
Fourth intercostal space at the left border of the sternum
Fifth rib between V2 and V4
Fifth intercostal space on the left midclavicular line
Left anterior axillary line at the horizontal level of V4
Left midaxillary line at the horizontal level of V4
- 37 -
Operation Preparations
Alligator Clip