18. If the wireless AP technology is used, the minimum distance between the radiator and
the human body should be not less than 20cm. Use the supplied antenna only. There
should be no shield in or around the room where the wireless AP is used.
19. Fix attention on the examination to avoid missing important ECG waves.
20. SHOCK HAZARD - Don't connect non-medical electrical equipment, which has been
supplied as a part of the system, directly to the wall outlet when the non-medical
equipment is intended to be supplied by a multiple portable socket-outlet with an
isolation transformer.
21. SHOCK HAZARD - Don't connect electrical equipment, which has not been supplied
as a part of the system, to the multiple portable socket-outlets supplying the system.
22. Do not connect any equipment or accessories that are not approved by the
manufacturer or that are not IEC/EN 60601-1-1 approved to the electrocardiograph.
The operation or use of non-approved equipment or accessories with the
electrocardiograph is not tested or supported, and electrocardiograph operation and
safety are not guaranteed.
23. Any non-medical equipment (such as the external printer) is not allowed to be used
within the patient vicinity (1.5m/6ft.).
24. Do not exceed the maximum permitted load when using multiple portable
socket-outlets to supply the system.
25. Multiple portable socket-outlets shall not be placed on the floor.
26. Do not use the additional multiple portable socket-outlet or extension cord in
the medical electrical system, unless it's specified as part of the system by
manufacturer. And the multiple portable socket-outlets provided with the system shall
only be used for supplying power to equipment which is intended to form part of the
system.
27. Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing
equipment and IEC/EN 60601-1 for medical equipment). Furthermore all
configurations shall comply with the valid version of the standard IEC/EN 60601-1-1.
Therefore anybody, who connects additional equipment to the signal input or output
connector to configure a medical system, must make sure that it complies with the
requirements of the valid version of the system standard IEC/EN 60601-1-1. If in
doubt, consult our technical service department or your local distributor.
WARNING
- 3 -
Safety Guidance