SE-601 Series Electrocardiograph User Manual
Q4: I want to print the hospital name in the report, but I can't find the place to enter it, where is
it?
A4: Please open the More Setup window, and move the cursor to the Institution textbox, and
then input the hospital name. The content you input in this textbox will be printed in the
report. For details, please refer to Section 9.9.4, "Entering Institution".
Q5: I pressed the START/STOP key, but the ECG didn't start printing, what's wrong with it?
A5: Please check whether there is any error information displayed on the screen.
If the hint No Paper or Paper Error is shown on the screen, please deal with it according to
the above-mentioned measures.
If the hint Transmitting... is shown on the screen, which means that the ECG is transmitting
the data to the PC, please wait a few seconds. You can start the printing after the data is
transmitted.
If none of the above-mentioned information is shown on the screen, please check whether
Sample Mode is set to Pre-Sample in the Work Mode Setup window. If it is so, the system
will not respond to the START/STOP key unless 10s data has been collected. In this case, all
you have to do is to wait a few seconds, and then you are able to start the printing by pressing
the START/STOP key again.
If the problem still exists, please contact the manufacturer or the local distributor for further
disposal.
Q6: I set the filter, speed and gain on the main screen1, but these settings were changed after
printing.
A6: The filter, speed and gain which are set on the main screen1 will not be saved, and they are
changed when you exit the main screen1 or after printing. If you want to save these settings,
please set them in the Record Setup window and the Filter Setup window.
3. Transmitting Problems
Q1: The ECG doesn't respond to any keys after a long time of transmission. It transmits nothing
for there is no new data appearing in the window of the PC software. What should I do?
A1: Some error may occur during the transmission course, for example, the connection between
the ECG and the net cable may loosen. In this case, please restart the ECG. If it doesn't work,
please restart the PC.
If the problem still exists, please contact the manufacturer or the local distributor for further
disposal.